UA Suite Calibration

Thank you for purchasing DEKIST UA products. The UA family is a transmitter that is inserted into the USB port and the sub-models are UA10, UA13 (UA30), UA20, UA50, UA51, A52, UA53, UA54, UA59, etc.

1.Connecting to a PC using a USB cable

  1. UA products PCConnect with, PCUA-Calibrator program 실행합니다.

* Download the program
1. UA Family USB Windows Driver
2. UA-Calibrator Program Download

2.UA 1x

The UA 1x consists of four sub-models. UA3 is Temperaturerelative humidityare measured simultaneously. UA11-K and UA11-T are thermocouple K type (High temperature environment) and T type (Low temperature environment) are supported respectively. UA13 uses a PT100 sensor Cryogenic Used for measurement.

2-1. UA10 (Temp/RH)

  1. In case of error of temperature value Full Range Input Offset You can use to change the value. For example, current temperature 24 ℃Full Range Input Offset 3 ℃ After entering and saving, the current temperature 3 ℃As high as 27 ℃Is displayed.

2-2. UA11 (K/T-Type)

1. Calibration button Limited useIt's possible. This is a function that can be used only by the manufacturer because specialized equipment is required for calibration work. User's Random useWhen a problem occurs due to A/S is paidProceeds to

2. In case of error of temperature value Full Range Input Offset You can use to change the value. For example, current temperature 25 ℃Full Range Input Offset -2 ℃ After entering and saving, the current temperature -2 ℃Temperature 23 ℃Is displayed.

3. responsivenessIn the case of Default value is 1Is designated as 1 ~ 15 It can be adjusted in units.
Sensitivity value If low The characteristics of the sensor Become sensitive Sensitivity value If high The characteristics of the sensor Dullly Works.

2-3. UA13 (PT100)

1. Calibration button Limited useIt's possible. This is a function that can be used only by the manufacturer because specialized equipment is required for calibration work. User's Random useWhen a problem occurs due to A/S is paidProceeds to

2. In case of error of temperature value Full Range Input Offset You can use to change the value. For example, current temperature 25 ℃Full Range Input Offset -2 ℃ After entering and saving, the current temperature -2 ℃Temperature 23 ℃Is displayed.

3.UA 50

The UA50 product performs monitoring to detect Total Volatile Organic Compounds using a metal oxide (MetalDxide) sensor.

The UA50 model adopts a separate metal oxide method. Therefore, no individual calibration gas operation is required and Zero CalibrationJust run it and use it. Depending on the desired usage environment, separate Zero CalibrationYou may need this. The zero point is basically adjusted at the time of product purchase and shipped.

CH1 : TVOC , CH2 CO2eq measurement screen
CH1 : TVOC , CH2 : Absolute humidity measurement screen

1. When there is an error in TVOC value TVOC offset You can change the value by using. For example, the current TVOC value is 0.5 ppmTVOC offset when -0.2ppm After entering and saving, the current TVOC value -0.2ppmThis down TVOC 0.3ppm The value is displayed.

2. responsivenessIn the case of Default value is 7Is designated as 1 ~ 15 It can be adjusted in units.
If the sensitivity value is low, the characteristics of the sensor Become sensitive Sensitivity value If high The characteristics of the sensor Dullly Works.

3. CH2 CO2eq OR Absolute humidity is Zero CalibrationIn the case of offset을 이용 하여 Error valueIt can give a change of and sensitivity cannot be adjusted.

3-1. Zero Calibration

UA50 connected to an auxiliary battery

1. Zero CalibrationFor Power supply onlyBecause it should be Auxiliary battery Connect to or USB HUB powerIt is recommended to put only.

2. Clean air, non-polluted, ventilated placeZero Calibration in 1 to 4 hours minimumI need this Recommended timeIs 4 hours.

3. Healing time 4 시간When this elapses, the product LED Red green colorto FlashingAnd CH1: TVOC, CH2: CO2eq value Zero Calibration Adjustment 완료It's possible.

4.UA 59

UA59-CO2-100-U product is designed to be used inside all chambers because it has a waterproof function on the stainless steel exterior with thermal conductivity sensor. A nest or nest mobile program is required to record measurement data.

  1. The screen consists of three columns. The upper part displays information about the product, communication status, and initialization. At the bottom, there are graphs for CO3 and temperature and buttons for changing the set value.
  2. Factory default is 'up to 25% BG gas N2', 5% Calibration.
  3. Carbon dioxide and the contained background gas are separated into nitrogen and air, and carbon dioxide concentration measurement limits can be set at 25% and 100%. In addition, in environments where the oxygen in the air is less than 18%, the carbon dioxide concentration limits can be set at 25% and 100%.
  4. Temperature is a value measured by a temperature sensor inside the product.
Updated on January 2022, 5